Storage capacity = 140m3

ICH-VICH, WHO, ASEAN conditions for climatic zones I, II, III and IV

  • 25 ± 2°C and 60 ± 5% RH: 3 chambers (16, 20 & 34 m³)
  • 30 ± 2°C and 65 ± 5% RH: 1 chamber of 21 m³ + 4 chambers of 1300 L
  • 30 ± 2°C and 75 ± 5% RH: 1 chamber of 20 m³
  • 40 ± 2°C and 75 ± 5% RH: 1 chamber of 12 m³
  • span lang="en-US">25 ± 2°C and 40 ± 5% RH: 1 chamber of 2000 L
  • 40 ± 2°C and <25% RH: 1 chamber of 2000 L 
  • 50 ± 2°C: 1 chamber of 100 L and 1 safety chamber
  • 5 ± 3°C: 1 chamber of 5 m³ (doubled regulation system)
  • -20 ± 5°C: 1 chamber

Photostability conditions (ICH Q1B & VICH GL5)

  • 1 cabinet of 1300 L
  • Visible : 1,2 millions lux.hours
  • UV : 200 watts.hours/m²
  • 25 ± 2°C

Other conditions

  • -80°C: 1 deep-freezer

Security of samples

  • 24/7 recording and monitoring system
  • Back-up chambers
  • Power generating unit
  • Several independent security levels
  • Automatic system of sample protection before any human intervention
  • Internal maintenance and metrology team
  • On-site storage of spare parts